Clinical Evaluation Strategy for a Women’s Health Device

Medical device clinical investigation for women health and CE marking support under EU MDR by a clinical research consultancy

A medical device start-up came to us with an active device for women’s health. They had already tested the device in more than 40 women through clinical investigations conducted in the US. They were preparing for a European launch. On paper, the remaining step looked simple: get the CE mark and start selling in Europe.

But they had not established the clinical evaluation strategy needed to support its European market entry. The EU Medical Device Regulation (MDR) expects a clinical evaluation plan (CEP) to come before the clinical evaluation report (CER), and they have never written such a plan.

The Clinical Evaluation Plan Was Missing

A clinical evaluation plan defines the strategy for collecting and evaluating the clinical evidence supporting the safety and performance of the medical device. Annex XIV of the MDR requires one for every device. It sets the evidence level proportionate to the device’s risk class and clinical claims.

A clinical evaluation plan must cover at least:

  • The device intended purpose and risk class, which sets how much evidence is needed
  • The general safety and performance requirements that need clinical data to support them
  • The target population groups, with clear indications and contraindications
  • The intended clinical benefits, with the specific endpoints
  • The methods for assessing clinical safety and side effects
  • The parameters for judging whether the benefit-risk ratio is acceptable, based on the state of the art in the corresponding medical field
  • How the evidence will be generated, through a systematic literature review, clinical investigations, or both, and how the sample size is justified
  • A clinical development plan from early studies to confirmatory investigations and post-market clinical follow-up

The mistake

Focusing on generating data without preparing the clinical evaluation strategy needed for European market entry. Regulatory pathways for medical devices differ significantly between Europe and the United States.

How We Fixed It: Establish the CEP Before Writing the CER

Building on the work carried out to date, we prepared:

  • Clinical evaluation plan (CEP): written for the program as it stood, without backdating it. A plan written after the investigations must say so, because the development record cannot change. A CE mark is still possible when the data and justification are sound.
  • State-of-the-art update: a systematic literature search of similar devices to set safety and performance benchmarks. But finding the relevant studies is only the first step. The evidence needs to be appraised, extracted, structured and analysed in a way that allows it to be used later to benchmark your own device.
  • Gap analysis: we identified and resolved inconsistencies across the technical documentation.
  • Clinical evaluation report (CER): the final document that brings all of this together for the notified body. Regulators expect a structured, evidence-based comparison that demonstrates how your device performs against the state-of-the-art and supports a clear conclusion about its safety, usability, and clinical performance.

The Result: CE Marking Obtained

The notified body review involved a few rounds of formal queries and written responses. We worked closely with the client to prepare well-supported responses to help move the review forward.

The device received its CE mark in 2026, a year and a half after submission. The company is now preparing its European launch.

Note

An 18-month review is common under the MDR, and long review times are one of the system’s most criticized features. In the EU, medical devices must undergo conformity assessment by notified bodies. A manufacturer can choose any of them across Europe, each one works differently, and none is bound by a fixed deadline.

Schedule Your Expert Hour

Get INSIGHTS into the strategies you need for a successful clinical trial!

Frequently Asked Questions

What is the difference between a clinical evaluation plan and a clinical evaluation report?

The plan comes first and sets out how the manufacturer will show the device is safe and performs as intended. The report comes last and presents the results: the clinical data, the literature review and the conclusion on the benefit-risk ratio. The notified body reads the report against the plan.

What is a state-of-the-art review in a clinical evaluation?

It is a structured review of the current standard of care and of similar devices already on the market. It sets the safety and performance benchmarks the new device is compared against, endpoint by endpoint.

What is the technical documentation for a medical device?

It is the full file a manufacturer submits to the notified body to show the device meets the MDR. It includes the device description, design and manufacturing information, risk management, the clinical evaluation and post-market surveillance plans. Every document has to be consistent with the others, because reviewers check them against each other.

Share This Post

More To Explore

Run your clinical trial in Europe Phase 1
Case study

Phase I Oncology Trial: Planning for Long-Term Data

A Nordic biotech company conducting a Phase I oncology trial in Austria had a protocol with conflicting pathways for long-term patient follow-up. We identified the inconsistency early, assessed its implications for budget, timelines and data collection, and amended the protocol within two weeks. The amendment was approved, and the collaboration extended to Phase II.

Medical device clinical investigation for women health and CE marking support under EU MDR by a clinical research consultancy
Case study

Clinical Evaluation Strategy for a Women’s Health Device

A women’s health medical device start-up had clinical data from more than 40 women but no clinical evaluation plan for European market entry. We addressed the gaps in its clinical documentation, established an EU MDR-aligned strategy and supported the company in obtaining CE marking.

Dr. Nora Gedeon

Data Protection Officer (DPO) and Vendor Manager

Nora holds degrees in law from Janus Pannonius University and in pharmacy from Semmelweis University, combining legal and scientific expertise. She is a certified Data Protection Officer with hands-on experience supporting GDPR compliance and contractual safeguards for biotech and pharma clients. With over 20 years in clinical research, she brings a pragmatic, risk-aware approach to data protection in global trials